CAIRE FreeStyle Comfort Technical Manual page 29

Hide thumbs Also See for FreeStyle Comfort:
Table of Contents

Advertisement

0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
High
Sieve Bed
Pressure
Failure
Feed/Waste
Valve Failure
Other
No Breath
Outlet Filter
Detected
Other
Low Battery
Battery Level
Shutdown
below 5%
Defective
Battery
O
Warmup
The unit is
2
Warming up
Low O
Intake Filter
2
Clogged
Sieve Bed
Failure
Low pressure
High pressure
Other
OCD
Kink in
Failure
Cannula
Outlet/Product
Filter Clogged
Other
Low Battery
Battery Level
Warning
below 10%.
Defective
Battery
FreeStyle Comfort
Replace sieve beds.
Replace feed/waste valves.
Contact Caire to send the
unit in for repair.
Check the cannula is leak-
free
Replace outlet filter.
Contact Caire to send the
unit in for repair.
Test unit with a charged
battery.
Test unit with another
battery.
Allow the unit to warm up
for 15 minutes.
Clean or replace intake
filters.
Replace sieve beds.
See low pressure
troubleshooting.
See high pressure
troubleshooting.
Contact Caire to send the
unit in for repair.
Ensure there are no kinks in
the cannula.
Replace outlet/product filter.
Contact Caire to send the
unit in for repair.
Test unit with a charged
battery.
Test unit with another
battery.
29