10.3 REINIT page
The initialization is the operation that restores the readings to zero.
1: Initialization time
Remember: it takes at least 40 minutes for the probe to warm-up. Before that time is reached,
you need to initialize the probe every minute or before each reading. After the probe is ready, it is
possible to set the time between two initializations. The default time is one minute. Otherwise,
you could also set the time for either 5, 10 or 15 minutes. The new time set will be activated after
the next initialization.
*** You will not be able to take measurements if you do not initialize the probe 30 seconds after
the software asks you to do it.
2: Active screen
You must select the Active Screen box. This function enables the touchscreen function of the
2
Archer
field computer. It is required to initialize the probe when the software asks it.
Instrumentation GDD Inc.
1
2
2018-01-16
24