#7 Accept the calibration
Press the MENU SELECT
down key to accept the
new value
pH Calibration
The PHS10 pH sensor was calibrated with pH 7.0 and pH 10.0 buffers at the factory before shipping.
Verify the displayed pH agrees with the actual pH. If the pH value deviates by more than 0.2 pH then
adjust the value in the Buffer: #1 pH: Menu by setting the existing pH 7 value to the actual sample value
and then accepting the calibration. For a detailed description see Electrode Standardization in the C‐22
manual.
pH Calibration Screen Shots
Calibration 2 point
(Zero pt. and Slope)
# 1 Home Screen
Press Menu Select
Up button
#3 Selection Menu
Set Up Screen
Press Menu Select
Up button
#5 Zero Point Calibration
Initiate the calibration by
simultaneously pressing
the horizontal CALIBRATE
keys with the sensor in a
7.00 pH solution, or...
#8 Return to Home
Screen
Press the CALIBRATE
down key to return to the
Home Screen.
#2 Selection Menu
Status Screen
Press Menu Select
Up button
#4 Selection Menu
Buffer Screen
(calibration)
Select BUFFER by
simultaneously pressing
the horizontal CALIBRATE
keys.
#5a Standardize to the
Sample pH value
Adjust the displayed value
to agree with the sample
value using the CALIBRATE
keys
Page 9
FCA‐22 Instruction Manual