Ph Sensor - ECD FCA-22 Instruction Manual

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
#7 Accept the calibration 
Press the MENU SELECT 
down key to accept the 
new value 
 
 
 
pH Calibration 
 
The  PHS10  pH  sensor  was  calibrated  with  pH  7.0  and  pH  10.0  buffers  at  the  factory  before  shipping.  
Verify the displayed pH agrees with the actual pH.  If the pH value deviates by more than 0.2 pH then 
adjust the value in the Buffer: #1 pH: Menu by setting the existing pH 7 value to the actual sample value 
and then accepting the calibration. For a detailed description see Electrode Standardization in the C‐22 
manual. 
 
pH Calibration Screen Shots 
 
Calibration 2 point 
(Zero pt. and Slope) 
# 1 Home Screen 
Press Menu Select 
Up button 
 
#3 Selection Menu 
Set Up Screen 
Press Menu Select 
Up button 
 
#5 Zero Point Calibration 
Initiate the calibration by 
simultaneously pressing 
the horizontal CALIBRATE 
keys with the sensor in a  
7.00 pH solution, or... 
 
   
 
 
 
#8 Return to Home 
Screen 
Press the CALIBRATE 
down key to return to the 
Home Screen. 
#2 Selection Menu  
Status Screen 
Press Menu Select 
Up button 
 
 
#4 Selection Menu 
Buffer Screen 
(calibration) 
Select BUFFER by 
simultaneously pressing 
the horizontal CALIBRATE 
 
keys.
 
#5a Standardize to the 
Sample pH value 
Adjust the displayed value 
to agree with the sample 
value using the CALIBRATE 
keys 
 
 
 
Page 9 
 
FCA‐22 Instruction Manual  
 
 
 