Product Specifications - ECD FCA-22 Instruction Manual

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
Product Specifications 
 
Sensors and Flow Train 
 
FCS (Free Chlorine Sensor) 
Polarographic electrode, Gold cathode and 
Silver anode, PTFE membrane 
 
PHD10 (pH sensor) 
Combined sensor, pH glass electrode, KCl/AgCl 
reference electrode with porous Teflon 
junction, 3 K‐ohm temperature sensor and 
signal conditioner 
 
Measurement Range 
 
Chlorine 
0.05 to 20 ppm 
pH  
 
4 ‐ 9 pH 
Temperature  0° to 50° C (32°‐ 122°F) 
Min/Max Flow  10 – 80 gal/hr (38‐300 l/hr) 
Pressure 
30 psig, max. (2 bar, max.) 
 
Wetted Materials 
 
PVC, Poly Propylene, PVDF, PTFE, Viton, Glass 
 
Process Connections 
 
Input 
¼" Barb fitting for tubing or ¼" 
FNPT with fitting removed 
Drain 
¾" FNPT 
 
Performance Characteristics 
 
Accuracy 
Response Time (T90) 
 
 
 
 
Drift 
 
Polarization time 
 
Electrolyte life 
 
   
 
Dependent on the 
accuracy of the DPD 
test and pH value 
Chlorine < 2 minutes 
Temp.  < 2 minutes 
pH < 30 seconds 
< 1.5% month 
First polarization, 60 
minutes 
Re‐polarization, 30 
minutes 
Up to 1 year  
 
 
C‐22 Analyzer 
 
Measurements 
Chlorine 
pH 
Temperature 
 
Cl
/pH Compensation  Between pH 5‐9 
2
 
Display 
Case  
Operating Conditions  Temperature ‐20° to 
Outputs 
Alarms 
 
 
Power 
Weight 
Shipping Weight 
 
 
 
Page 22
 
FCA‐22 Instruction Manual  
0.00 to 20 ppm 
0 – 14 pH 
0° to 100° C (32°‐212°F) 
2.5" X 1.75" backlit LCD 
4 lines of text and 
Graphical 
NEMA 4X, LxWxD 5.7" x 
5.7" x 7" 
70° C, Protect from 
direct sunlight and 
excessive heat. 
Humidity, RH from 5‐
95% non‐condensing 
One 4‐20mA for Free 
Chlorine 
Optionally up to 4, PID 
optional 
(2) Form C, SPDT 
230VAC 5A; 30 VDC 5A 
 
Optionally up to (8) 
110/220 VAC @ 50‐60 
Hz or optionally 24 VDC 
@ 0.25A  
11 lbs (5 kg) 
14 lbs (6.5 kg) 
 
 

Advertisement

Table of Contents
loading

Table of Contents