Troubleshooting; Troubleshooting Guide - ECD FCA-22 Instruction Manual

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
Troubleshooting 
 
The  FCA‐22  was  evaluated  and  calibrated  at  the  factory  before  shipment.  Upon  initial  start  up  the 
system should require minimal to no adjustments. 
Verify  the  system  has  adequate  flow,  greater  than  10  gals  /hr,  the  pH  electrode  and  the  temperature 
sensor are reading correctly. These parameters effect the measurement and must meet the standards 
listed  in  the  Calibration  Section  above.  If  these  conditions  are  met  and  problems  still  exist  use  the 
Troubleshooting Table to find a remedy. 
Troubleshooting Guide 
 
Symptom 
 
Displayed value is Higher 
than DPD test value. 
 
 
Displayed value is Lower than 
DPD test value 
 
 
 
 
 
 
 
   
 
 
 
Possible Cause 
 
Insufficient Polarization time 
 
 
Damaged Membrane Cap 
 
pH indication higher than 
actual value 
 
Temperature indication lower 
than actual value 
 
 
 
 
Electrical short or wet 
connection inside the sensor 
or cable assembly 
 
 
 
 
Chloramine or other oxidants 
present in sample yielding a 
high DPD test. 
 
Coated or dirty membrane 
 
 
 
 
 
Page 20
 
FCA‐22 Instruction Manual  
Remedy 
 
Polarize FCS for full 90 
minutes before calibration. 
 
Replace Membrane Cap 
 
Calibrate pH sensor or adjust 
value on uncompensated FCA 
 
Calibrate the Temperature 
(see Calibration) The 
temperature sensor lags the 
process temperature wait for 
temperature equilibrium. 
 
Remove measuring chamber 
and dry the cathode surface, if 
the indication does not go to 
zero there is leakage. Replace 
the sensor. 
 
 
 
 
 
Replace water and chlorinate 
 
 
 
Clean or replace the 
membrane 
 
 
 

Advertisement

Table of Contents
loading

Table of Contents