ECD FCA-22 Instruction Manual page 11

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
Perform a DPD test on the sample and enter the measured ppm Cl
and accept the calibration. The displayed Cal value, x.xx  mV/ppm Cl
For a detailed description see Electrode Span Calibration in the C‐22 manual.  
 
**Record the displayed Cl
the DPD test then correct the input value appropriately**. 
 
The  zero  point  of  the  sensor  is  extremely  stable  and  should  not  need  adjusting.  If  the  calibration  is 
needed then place the sensor in a sample of chlorine free process water and allow it to equilibrate for ½ 
hour. Initiate the standardization in the Buffer: #1 ppm Menu. The Standardize Value will be 0.00 ppm 
Cl
 and the Cal value will display the new mV signal from the sensor. 
2
 
Chlorine Calibration Screen Shots 
 
#1 Home Screen 
Press Menu Select 
 
Down button
#3 Selection Menu 
Select BUFFER by 
simultaneously pressing 
the horizontal CALIBRATE 
keys. 
 
#5 Adjusting zero point  
If a non‐zero ppm  
solution is used then  
adjust the value by 
selecting the digit with the 
horizontal keys and the  
value with the vertical key.
#7 Leave Zero Calibration  
Screen 
Press Menu Select 
 
Down button
   
 
 ppm reading at the sample time, if the value has changed before finishing 
2
 
Page 11
 
 
 value in the Buffer: #2 ppm Menu 
2
 will change to the new slope value. 
2
#2 Graphical Display 
Press Menu Select 
 
Down button
#4 Zero Point Calibration 
(optional step, set at factory,  
move to #7) 
Initiate the calibration by 
simultaneously pressing 
the horizontal CALIBRATE 
keys with the sensor in a  
0 ppm chlorine solution. 
#6 Accepting the 
Calibration 
Accept the calibration by 
pressing the Menu Select 
 
down key.
#8 Setting Chlorine Slope 
Initiate the calibration by 
simultaneously pressing 
the horizontal CALIBRATE 
keys with the sensor in a  
known chlorine solution.
 
 
FCA‐22 Instruction Manual  
 

Advertisement

Table of Contents
loading

Table of Contents