ECD FCA-22 Instruction Manual page 21

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
Displayed value is Lower than 
DPD test value (cont'd) 
 
Zero Chlorine Reading 
 
Unstable Chlorine Reading 
 
 
 
   
 
 
 
Low tension on the 
membrane 
 
 
 
Flow to low through the flow 
cell 
 
 
Air bubbles trapped on 
membrane 
 
 
Air bubble inside the sensor 
between cathode and 
membrane 
 
pH indication lower than 
actual value 
 
Temperature indication is 
higher than actual value. 
 
 
No electrolyte in the sensor 
 
 
Open Circuit on FCS, broken 
or bad electrical connection 
 
 
Air bubbles on the membrane 
 
 
 
Changing temperature, the 
lag of the temperature sensor 
looks like drift  
 
 
Page 21
FCA‐22 Instruction Manual  
Verify the Measuring 
Chamber is fully tightened 
onto the body or replace 
membrane. 
 
Clean CHFC, fittings and 
tubing, verify the sample feed 
rate at +10 gal/hr. 
 
Loosen FCS fitting and lift 
sensor slightly to purge air 
trapped in the flow cell. 
 
Refill sensor, see Maintenance
 
 
 
Calibrate pH sensor or adjust 
value on uncompensated FCA 
 
Calibrate the Temperature 
(see Calibration) The 
temperature sensor lags the 
process temperature wait for 
temperature equilibrium. 
 
 
Refill Sensor 
 
 
Check connector and wiring to 
the connector inside the C‐22 
 
Loosen FCS fitting and lift 
sensor slightly to purge air 
trapped in the flow cell. 
 
Wait for equilibrium 
 
 
 