#9 Setting the Chlorine
ppm
Use the horizontal
CALIBRATE key to select the
digit to be adjusted. Set the
screen value to the DPD test
reading.
#11 Adjusting the Value
Move to the next digit to be
adjusted with the horizontal
Calibrate keys.
#13 Accepting the
Calibration
Accept the calibration by
pressing the Menu Select
down key.
Commissioning Outputs
4‐20 mA Range
The FCA 22 4‐20 mA output is configured 0‐20 ppm Free Chlorine. The Range can be set to other values
to accommodate the user's requirement in the Set Up Menu. To access the Set Up menu scroll down
from the Home Screen using the down MENU SELECT key. Enter the menu by simultaneously pressing
the horizontal CALIBRATE keys and then scroll down to the 4 mA screen using the MENU SELECT key.
Adjust the value with the CALIBRATE keys and accept the change with the down MENU SELECT key.
Repeat for the 20 mA screen. For more details see the C‐22 Manual.
Setting the 4‐20 mA Range Screen Shots
#1 Home Screen
Press Menu Select
Down button
#10 Adjusting the Value
Adjust the numeric value
with the vertical CALIBRATE
keys. Adjust largest digit
first.
#12 Adjusting the Value
Adjust the numeric value
with the vertical CALIBRATE
keys.
Once the value has been
set the analyzer reading
will stabilize.
#14 Return to the Home
Screen
Leave the Calibration Screen
and return to the Home
Screen by pressing the
CALIBRATE down key.
#2 Graphical Display
Press Menu Select
Down button
Page 12
FCA‐22 Instruction Manual