Ma Output - ECD FCA-22 Instruction Manual

Free chlorine analyzer/controller
Table of Contents

Advertisement

 
 
#9 Setting the Chlorine 
ppm 
Use the horizontal 
CALIBRATE key to select the 
digit to be adjusted. Set the 
screen value to the DPD test 
reading. 
#11 Adjusting the Value 
Move to the next digit to be 
adjusted with the horizontal 
Calibrate keys. 
 
#13 Accepting the 
Calibration 
Accept the calibration by 
pressing the Menu Select 
 
down key.
 
 
Commissioning Outputs 
4‐20 mA Range 
The FCA 22 4‐20 mA output is configured 0‐20 ppm Free Chlorine. The Range can be set to other values 
to accommodate the user's requirement in the Set Up Menu. To access the Set Up menu scroll down 
from the Home Screen using the down MENU SELECT key. Enter the menu by simultaneously pressing 
the horizontal CALIBRATE keys and then scroll down to the 4 mA screen using the MENU SELECT key. 
Adjust the value with the CALIBRATE keys and accept the change with the down MENU SELECT key. 
Repeat for the 20 mA screen. For more details see the C‐22 Manual. 
Setting the 4‐20 mA Range Screen Shots 
#1 Home Screen 
Press Menu Select 
Down button 
 
   
 
 
 
#10 Adjusting the Value 
Adjust the numeric value 
with the vertical CALIBRATE 
keys. Adjust largest digit 
first. 
 
#12 Adjusting the Value 
Adjust the numeric value 
with the vertical CALIBRATE 
keys. 
Once the value has been  
set the analyzer reading  
 
will stabilize.
#14 Return to the Home 
Screen 
Leave the Calibration Screen 
and return to the Home 
Screen by pressing the 
CALIBRATE down key. 
#2 Graphical Display 
Press Menu Select 
Down button 
 
 
Page 12
 
FCA‐22 Instruction Manual  
 