Figure 35 - Daisy-Chain Configuration (Does Not Apply To Model 8600-2); Table 19 - Daisy -Chain Operation - Geokon 8600-2 Instruction Manual

Micro-6000 multi-channel datalogger
Table of Contents

Advertisement

34
SW1 Setting
2
3
4
OFF
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF
OFF
ON
ON
ON
OFF
OFF
ON
OFF
ON
ON
ON
OFF
ON
ON
ON
Figure 35 is a schematic representations of the daisy-chain configuration with multiplexers
configured for 32-channels. It shows three multiplexers sharing the same control ports, and a
single cable is used to interconnect them.
VW Gages 1-32
VW Gages 33-64
VW Gages 65-96

Figure 35 - Daisy-Chain Configuration (Does not apply to Model 8600-2)

1-32 (32-channel mode), 1-16 (16-channel mode) DEFAULT MUX1
33-64 (32-channel mode), 17-32 (16-channel mode) MUX2
65-96 (32-channel mode), 33-48 (16-channel mode) MUX3
97-128 (32-channel mode), 49-64 (16-channel mode) MUX4
129-160 (32-channel mode), 65-80 (16-channel mode) MUX5
161-192 (32-channel mode), 81-96 (16-channel mode) MUX6
193-224 (32-channel mode), 97-112 (16-channel mode) MUX7
225-256 (32-channel mode), 113-128 (16-channel mode) MUX8
Table 19 - Daisy-Chain Operation/Channels Accessed
Micro-6000 Datalogger
U5
Reset
Reset
IN
H
H
OUT
Reset
IN
H
H
OUT
Reset
IN
H
Channels Accessed
C4
Clock
Clock
Multiplexer #1
J
MUX1
J
SW1 SETTING:
2 3 4
OFF OFF OFF
Clock
Multiplexer #2
J
MUX2
J
SW1 SETTING:
2 3 4
OFF OFF ON
Clock
Multiplexer #3
J
MUX3
SW1 SETTING:
2 3 4
OFF ON OFF

Advertisement

Table of Contents
loading

This manual is also suitable for:

8600-1

Table of Contents