F.4 Troubleshooting
Below are some commonly experienced problems along with possible remedial action. Contact
the factory if any problem remains unresolved or additional help is required.
Symptom: A particular channel on the multiplexer appears to be malfunctioning:
 Check sensor connections on the terminal board. Clean if corrosion exists.
 Try moving the sensor wired to the suspect channel to another channel to verify the
malfunctioning of the channel (as opposed to the sensor).
Symptom: No channels are working:
 Inspect circuit board for shorts, opens, or other damage.
 Is moisture present on circuit board? If so, install desiccant to absorb.
Symptom: Channel selection appears to be random:
 Has corrosion built up on the circuit board? Clean if necessary.
 Is there a source of electrical noise nearby? Move multiplexer or noise source if possible.
F.5 8032 Multiplexer Specifications
GENERAL:
Power Requirements
Quiescent Current
(MICRO-800/MICRO-1000 mode)
Quiescent Current (GK-403 mode)
Channel Activated Current
Control Line Input Impedance
Control Line Input Levels
Transient Protection
Operating Temperature
RELAYS:
Type
Power
Contact Type
On Resistance
Coil Resistance
Maximum Switching Power
Maximum Switching Voltage
Maximum Switching Current
Operate Time
10-16 VDC (unregulated)
80 µA (16CH mode)
130µA (32CH mode)
12 mA
30 mA
100 kΩ (CLOCK), 100 kΩ (RESET)
TTL or RS-232 (±9 VDC)
16 VDC Transzorbs
−40 to +60° C
NAIS TXS2SA-4.5V DPDT non-latching
11.1 mA @ 5VDC (55.5 mW)
Gold clad silver alloy
100 mΩ
405 Ω
30W (resistive)
110 VDC
1 A
≈5 msec Max.