6
Symptom: The Datalogger will not operate on external power:
The external voltage supply is below operating limits. If the external source is a battery,
charge it. If it is a power supply, check the output with a voltmeter.
Verify both Load and Charger Fuses on Power Distribution Board (Refer to Section 4.5).
Symptom: Loss of Micro-6000 program and/or data:
The system has experienced a voltage dropout or surge which disrupted operations.
The surge originated as a result of lightning. Install appropriate grounding. Install
lightning protection devices on all incoming and outgoing lines (consult factory).
Verify .CR6 Program is running on the Datalogger (See Appendix E).
Symptom: Sensor readings show "NAN" or are unstable:
The wrong sensor type has been connected. Check the model number of the sensor
against the requested sensor list.
The cable(s) to the sensor(s) have been damaged permitting moisture and debris to enter
the jacket. Wires may be shorted together. Inspect the cable and perform Ohm test to check for
short circuits.
If all sensors on a particular Multiplexer are erratic or returning "NAN" perhaps the
Multiplexer or cable has been damaged. Inspect the cable. Try another cable to verify proper
functioning.
The sensor(s) have been damaged. For example, overrange on a vibrating wire sensor can
cause erratic readings.
There is an electrical noise source nearby. Move the sensor, cables, and Datalogger away
from the noise source. Install grounding devices. Consult the factory.