Connector And Cable Wiring - Geokon 8600-2 Instruction Manual

Micro-6000 multi-channel datalogger
Table of Contents

Advertisement

30
F.3.2 Connector and Cable Wiring
Inside
10 Pin
J4
Color
Bendix
1
Brown
2
Red
3
Orange
4
Yellow
5
Green
6
Blue
7
Purple
8
Grey
9
White
10
P1 Terminal
1,2
3,4
5,6
7,8
9,10
11,12
13,14
15,16
17,18
19,20
21,22
23,24
Note: P1 is a "stacking" type connector. Odd number terminals are read left to right on the
bottom row. Even numbered terminals are read left to right on the top row.
Description
A
COM HI 1
B
COM LO 1
C
COM HI 2 ( 16-channel)
D
COM LO 2 ( 16-channel)
K
Analog Ground
F
+12 Volt Power
G
Power Ground
RESET (DATALOGGER)
H
SENSE (GK-403)
J
CLOCK
E
No Connection
Table 14 - J4 (I/O) Connector
Label
SHLD
SHLD
C1H
C1L
C2H
C2L
SHLD
+12V
GND
RESET
CLOCK
SHLD
Table 15 - P1 (I/O) Connector
8032-5 (TAN) Cable
Wire Color
White
White's Black
Red
Red's Black
Shield Drain Wires – all pairs plus
overall
Yellow
Yellow's Black
Green
Green's Black
Blue and Blue's Black (unused)
Description
SHIELD
SHIELD
COM HI 1
COM LO 1
COM HI 2 ( 16-channel)
COM LO 2 ( 16-channel)
SHIELD
+12 Volt Power
Power Ground
RESET
CLOCK
SHIELD

Advertisement

Table of Contents
loading

This manual is also suitable for:

8600-1

Table of Contents