USER MAINTENANCE
7
Compare the reference device reading with the chamber
temperature display again.
If the reference device and the chamber temperature
•
display readings are the same or the difference falls
within the range of your protocol, the incubator is now
calibrated for temperature.
See the next step if the readings fail to match or fall
•
outside of your protocol range.
8
If the two readings are not the same, and the difference still
falls outside the acceptable range of your protocol, repeat
steps 3 – 7 up to two more times.
Three calibration attempts may be required to successfully
calibrate units that are more than ±2°C out of calibration.
9
If the temperature readings of the incubator temperature display and the reference device
still fall outside your protocol after three calibration attempts, contact your incubator
distributor or
Technical Support
- OR -
for assistance.
End of procedure