USER MAINTENANCE
5) Temperature Stabilization
The incubator air temperature must stabilize in order to perform an accurate calibration.
Allow the incubator to operate undisturbed with the doors shut for at least 8 hours when
•
first putting the unit into operation in a new environment.
To be considered stabilized, the incubator chamber must operate at your calibration
•
temperature for at least 1 hour with no fluctuations of ±0.1°C or greater.
Start
Required temperature stabalization period operating undisturbed with the incubator door(s) closed.
Suggested Temperature Calibration
1
Once the incubator temperature has stabilized, compare the
reference device and incubator temperature display readings.
If the readings are the same, or the difference between the
•
two falls within the acceptable range of your protocol, the
display is accurately showing the incubator air
temperature. The Temperature Calibration procedure is
now complete.
If a difference falls outside of your protocol range, advance
•
to step 2.
2
A display calibration adjustment must be entered to match the
display to the reference device. See next step.
Fluctuations
(Exaggerated)
- Or -
Continued next page
Begin Calibraiton
Reference Device
SET TEMPERATURE
Reference Device
29 |
P a g e