Operation
3.3.5
**********************
PROGRAM NO. 1
INFO MEMORY
*********************
3.3.6
**********************
PROGRAM NO. 1
BOTTLE MEMORY
**********************
3.3.7
**********************
INPUT SIGNALS
FILLING LEVEL 100
FLOW ANAL. 12.2 mA
EVENT 0 / FLOW DIG 0
**********************
3.3.8
**********************
INTERNAL
TEMPERATURE
3.6 °C
*STORED VALUES
**********************
3.3.9
**********************
PROGRAM NO. 1
ANALOGUE FLOW
SIGNAL
**********************
3.3.10
**********************
PROGRAM NO. 1
MANUAL SAMPLE
**********************
4.6.1
**********************
MANUAL SAMPLE:
BOTTLE 1
**********************
ba75817e01
05/2009
By pressing the ENT key, the menu „INFO MEMORY" is displayed.
Here you can have a look at the stored messages.
By pressing the ENT key the bottle memory is displayed as follows:
- Number of the bottle
- Date and filling period (from – to)
- Total filling time
- Taken and requested samples.
With the up/down keys you can scroll to each bottle.
Display of „Input Signals":
-
Filling level (state of level electrodes. The indicated value has no
dimension!)
-
Flow analogue
-
Event input
-
Flow digital
Display of the temperature inside the sample compartment.
By pressing the ENT key, the stored mA-signal is shown. The
values displayed are hourly average values.
Here you can trigger manually a single sample extraction into any
bottle (bottle No. can be selected).
After having pressed the ENT key, you can either confirm the bottle
number already displayed or enter another bottle number. If a
different bottle number is entered, the distributor will change to the
previous bottle after the sample extraction.
This sample is stored in the memory as „manual sample".