3. INSTALLATION

Wall Mounting

For flush mounting (chased-in cables), refer to the electrician's wiring supplement
supplied with Electrician's Kit DRK4064
CAUTION: This product must be earthed.
Remove the 4 cover screws with the hex
wrench provided
Release the earth bond from the back-
plate by removing the nut and
shake-proof washer
Using the back-plate as a template,
mark its position on the wall, bearing in
mind the final position of the light-head
Fix the bracket to the wall
Whilst supporting the weight of the
power supply module and
light, replace the earth
bond to the back-plate
with the shake-proof
washer and nut
Fit the power supply
module to the wall
bracket with the 4 cover
screws
Connect the IEC flex to
the socket on the side of
the power supply module
Plug the IEC flex into a
13A mains socket and
check for correct
operation

Rail mounting

Hold the rail against the selected location
on the wall, bearing in mind the final
position and range of travel of the light-
head required to project its light patch
where it will be needed.
Ensuring the rail is horizontal; mark the
positions for the 2 screw holes.
Fix the rail to the wall, using appropriate
plugs as necessary and the 2 screws
provided.
NOTE: Slide the screw-head cap retainers the
correct way up onto the screws before fitting
the screws