DARAY X740 Operating & Installation Manual

DARAY X740 Operating & Installation Manual

Led examination light

Advertisement

Quick Links

X740
LED Examination Light
Operating & Installation Manual
QAM.X740.0519.4

Advertisement

Table of Contents
loading

Summary of Contents for DARAY X740

  • Page 1 X740 LED Examination Light Operating & Installation Manual QAM.X740.0519.4...
  • Page 2 Page 1 of 30...
  • Page 3: Table Of Contents

    Table of Contents 1. Introduction ................3 1.1 Supported Lights ..............4 2. Technical Specifications ............5 2.1 Pre-Installation Responsibilities and component packing list ... 6 2.2 Pre-Start Checks ..............7 3. Operation ................8 3.1 Range of Motion ..............8 3.2 Powering On &...
  • Page 4: Introduction

    DARAY. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by DARAY for the use or reliability of software or equipment that is not supplied by DARAY.
  • Page 5: Supported Lights

    X740 LED Mobile Weighted Base Light X740LE1 X740 LED Hardwired W3 BESA Wall Mounted Light X740LE2 X740 LED Hardwired W4 Spreader Plate Wall Mounted Light X740LE3 X740 LED Hardwired W5 Panel Mounted Light X740LE4 X740 LED Hardwired W6 Wall Mounted w/ Integral PSU...
  • Page 6: Technical Specifications

    2. Technical Specifications Light Source Light Intensity (Lux) @ 0.5m 80,000 Lux ±5% Light Intensity (Lux) @ 0.75m 40,000 Lux ±5% Light Intensity (Lux) @ 1.0m 20,000 Lux ±5% Light Intensity Adjustment Range 25% - 100% Colour Temperature (Kelvin) 3,600 – 5,000K ±500 Colour Rendering Index (CRI) Colour Rendering (R9) @ 0.5m: 60mm...
  • Page 7: Pre-Installation Responsibilities And Component Packing List

    2.1 Pre-Installation Responsibilities and component packing list This document is a guide to the steps that need to be performed to correctly install the X740 Examination light. However, the work required to be performed is the responsibility of the owner or designated contractor/s.
  • Page 8: Pre-Start Checks

    X740LE3 Qty. X740LE4 12V Plug top transformer Light head and spring rigid arm system incl. base box (transformer integrated) Panel Mount Bracket (W5) Microfibre cloth Light head and rigid arm system User Manual Microfibre cloth User Manual 2.2 Pre-Start Checks Please check that the power cable is without visible damage, harsh folds, knots or cuts in the rubberised coating.
  • Page 9: Operation

    3. Operation 3.1 Range of Motion The full range of motion between the X740 Range varies. This is an image depicting the range of motion on a standard wall mount fitting. Page 8 of 30...
  • Page 10: Powering On & Variable Intensity

    3.2 Powering On & Variable Intensity The X740 uses a control panel on the side of the light-head to power the light on and off, and to adjust the brightness (intensity), and colour temperature. Press the power button to power ON or OFF.
  • Page 11: Adjusting The Light Head

    3.3 Adjusting the Light Head The X740 head is secured using an adjustment knob on the side of the arm, near the head. This also controls the forward and backward motion of the head itself. Thumb Screw Loosening this thumb screw will allow you to position the light-head up or down.
  • Page 12: Installation

    4. Installation 4.1 Wall mounted versions Compatible versions: X740LW / X740LE1 / X740LE2 / X740LE3 / X740LE4 The wall bracket needs to be true on the vertical plane. PIVOTW2: supplied with PIVOTW3: supplied X740LW with X740LE1 PIVOTW4: supplied with X740LE2 PIVOTW5: supplied Page 11 of 30 with X740LE3...
  • Page 13: Considerations

    It may be necessary to adjust this height, if there are specific medical procedures the light will be used for. If in doubt consult the end user or DARAY. When installing the bracket take consideration of where you want the light head to sit when in its lowest position.
  • Page 14: X740Lw - Placement And Installation

    UK wall socket. Please note, under no circumstances should this product be connected directly to the Mains electrical supply, this product requires a Low Voltage Connection only and must only be used with a Daray Certified Power Adapter. Page 13 of 30...
  • Page 15 The light can now be tested by connecting the DC jacks and applying power. Press the switch on the head to power your light on. See page 9 Check arm and head for smooth and correct movement. Your new DARAY light is now ready for use.
  • Page 16: X740Le1 - Placement And Installation

    4.1.3 X740LE1 – Placement and Installation  Remove the light from its packaging.  This version includes our PIVOTW3 BESA mounting bracket, for mounting to a BESA box. Ensure the BESA back-box is mounted between recommended 850- 1100mm from the floor to the underside of the bracket – dependent on couch height.
  • Page 17: X740Le3 - Placement And Installation

    4.1.5 X740LE3 – Placement and Installation  Remove the light from its packaging.  This version includes our PIVOTW5 panel mounting bracket.  Fit front bracket and rear plate to pre-drilled panel.  Fit light spigot into bracket and pass cable through to the rear of panel.
  • Page 18 CAUTION: This product must be earthed. Check you have all of the items included in the checklist on Page 7 Note the mounting position advice and warnings offered in Section 4.1.1 Considerations on page 10 Flush-fitting the wall box (chased-in cables): ...
  • Page 19: Rail Mounted Version (X740Lr)

    4.2 Rail mounted version (X740LR)  Remove the light from its packaging.  If you have also purchased our XRAIL accessory, hold the rail against the selected location on the wall, bearing in mind the final position and range of travel of the light-head required to project its light patch where it will be needed.
  • Page 20: Mobile Version (X740Lm)

     Mark the position on the desk for mounting. Use screws the fix the bracket securely.  Fit the spigot on the end of the light arm into the top of the clamp, make sure it is fully secured in place. Apply the power cable, and power the light on.
  • Page 21  Insert the orange IEC cable into the black power supply attached to the base upstand.  The light can now be tested by connecting the DC jacks between the power supply and the light, and applying power. Press the switch on the head to power your light on.
  • Page 22: Optional Battery Back-Up

    4.5 Optional battery back-up 4.5.1 Packing List  Battery back-up unit  Double-ended bridge cable  12V PSU + extension cable 4.5.2 Install Plug the power adapter into the wall and insert the DC connector into the battery. Switch the battery to the ON position and the red light on the top of the battery pack will illuminate.
  • Page 23: Maintenance

    The proposed maintenance is only a suggestion. Depending on the use of the product and the operating environment, this may need to be revised more often. DO NOT ATTEMPT ANY FIXES BEFORE CONSULTING DARAY. If you cannot resolve a problem then please contact our helpline on 0800 878 9864 or email support@daray.co.uk...
  • Page 24: Product Cleaning & Care Guidelines

    Acetone-based cleaners (e.g. nail polish remover, Goo-off™)  MEK (Methyl Ethyl Ketone) If you are unsure whether you are using the correct cleaning fluids on your DARAY equipment, please feel free to get in touch with us for advice. Page 23 of 30...
  • Page 25: Head Adjustment

    5.4 Head Adjustment The head of the X740 light has a 180 degree forward and backward motion; this is adjusted secured by the thumb screw at the bottom of the light head. The head has a total turning capability of 320 degrees, attempting to push further may cause damage to the movement and internal wiring of the light.
  • Page 26: Rigid Arm Adjustments

    5.6 Rigid Arm Adjustments The X740 light is mounted on spring-tensioned rigid arms which are balanced specifically to be able to be moved using a single finger touch. Your light will have been balanced during manufacture, so there should be no need for you to make any adjustments out of the box.
  • Page 27: Troubleshooting Guide & Spare Parts

    6. Troubleshooting Guide & Spare Parts Problem Possible Cause Corrective Actions Wall Version - Check the plug top adapter is plugged in and switched Power Supply on at the power socket. Press the button on the back of the On/Off/Intensity light to check if the light will power Button on (if dimmed refer to poor light...
  • Page 28 12V PSU and 3m IEC Cable Kit DRK4176 (mobile) For any SPARE PARTS or TECHNICAL ISSUES phone DARAY’S Service Delivery Team on: +44 (0) 800 878 9864 Opening times 9am–5pm Monday to Thursday (excluding bank holidays) 9am-2:30pm on Friday Or email: support@daray.co.uk...
  • Page 29: Warranty Information

    DARAY Ltd, please read the following information carefully. The DARAY Ltd returns policy provides guidance on when you can return goods we have supplied, and what you can expect from us once you do. To see our detailed returns policy and procedure visit www.daray.co.uk/returns...
  • Page 30: Warranty Details

    Direct damage to your property caused by the proven negligence of Daray. 6. This agreement does not give any rights other than those expressly set out above and in particular, Daray will not be responsible for any loss of income, profits or contracts or any direct or indirect consequential loss, damage caused to or suffered by the purchaser as a direct result of this agreement.
  • Page 31 Service Notes Page 30 of 30...
  • Page 32 DARAY Ltd. Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH  +44 (0)800 878 9864  +44 (0)333 321 0973 sales@daray.co.uk  www.daray.co.uk...

Table of Contents