2.4 Mains Voltage Strapping

This unit may be field-strapped to any of the rated mains voltages shown in Section 2.2 of this manual. Use
the following procedure:
1. Unsolder the zero-ohm resistors from the A2 PWB, JP1 through JP5 and reinstall them for the
desired mains rating according to Table 1 in Figure 5.8. Insulated, solid hookup-wire jumpers
may be used instead of zero-ohm resistors.
2. Replace mains fuses A2F1 and A2F2 per Table 2, Figure 5.8.
3. Use an indelible marking pen to alter the mains voltage and current ratings on the nameplate
according to Section 2.2 specifications. DO NOT alter the model number or serial number.
4. Reassemble the unit, connect it to the new mains voltage source and perform Mains Leakage and
RF Power Output testing according to Section 4.2 of this manual.

2.5 Environmental Protection

At the end of its service life, the Hyfrecator® 2000 should be disposed of locally in accordance with local
regulations. Component materials are:
• Thermoplastic enclosure and stainless steel mounting plate.
• Thermoset printed wiring boards containing miscellaneous electronic components.
• Power transformer made of steel and copper.
• Mains cord and pencil accessory made of thermoplastic and copper.
• Accessory electrodes are stainless steel and thermoplastic.
• Electrodes contaminated with biological waste should be disposed of as biologically hazardous material.
• Shipping container and packing material are a combination of cardboard and plastic film.