Figure 25. Ms2Wfp Outline Drawing And Mains Voltage Selection - Bartington MS2 Operation Manual

Magnetic susceptibility system
Table of Contents

Advertisement

20.1.1. Mains Voltage Selection
6
5
Key to Figure 25
1. PC11
2. Heatsink
3. Voltage selector (110V or 240v)
4. Transformer
5. Controller & power supply
Before connecting to a mains electricity supply, check that the MS2WFP power supply is set to
the required AC supply voltage. The voltage selection is performed using the voltage selector
switch, which is mounted on the bottom left hand side of the rear panel of the power supply
unit. This is shown in Figure 26. The selected voltage should be displayed on a label positioned
by the power inlet socket. To alter the supply voltage, proceed as described in the figure, always
remembering to record the new voltage on the label.
BARTINGTON INSTRUMENTS
7
8
4
All signal lines down this side
All power lines down this side

Figure 25. MS2WFP outline drawing and mains voltage selection.

6. Ammeter
7. Capacitor C2
8. Rectifier
9. Capacitor C1
10. Main earth point
Page 62 of 82
9
10
3
1
2
OM0408/49