For older units, it will be necessary to remove the two-way plug from the replacement battery,
cutting each wire separately so as not to short the battery. Assemble the terminal block onto the
battery wires, then reconnect being careful to connect black to black and red to red.
Re-assemble top half of enclosure and secure with the appropriate screws or slotted nuts.
Dispose of the old battery in an approved disposal facility.

5.8. Data Transmission

Two-way transmission of data may be accomplished over a screened cable up to 50 metres.
A 2-metre cable with a 9-way 'D' type connector is provided.
The baud rate and bit format are set by a rotary switch on the rear panel as follows:
A. Standard serial interface
Baud rate
Bits per character
Number of characters
Character format
Control lines
Code
B. Serial interface for κ/T system operation
(Used for compatibility with the temperature meter)
Baud rate
Bits per character
Number of characters
Character format
Control lines
Code
C. Serial interface for 9600 baud option
Baud rate
Bits per character
Number of characters
Character format
Control lines
Code
BARTINGTON INSTRUMENTS
1200
1 start, 7 data, 2 stop, no parity
5 + carriage return
Sign, four digits, CR
Inactive
ASCII
1200
1 start, 8 data, 2 stop, no parity
5 + carriage return
Sign, four digits, CR
Inactive
ASCII
9600
1 start, 8 data, 2 stop, no parity
5 + carriage return
Sign, four digits, CR
Inactive
ASCII
Page 20 of 82
OM0408/49