Installation Instructions

Ambient Conditions

The Sartorius Basic Balance is designed to provide reliable weighing
results under normal conditions in the laboratory and in industry.
When choosing the area to Setup your balance, avoid exposing it to
the ambient conditions listed in the following so that you will be able to
work with added speed and accuracy:
- extreme heat radiation
- drafts
- inclined position
- extreme vibrations
- aggressive chemical atmospheres
The balance may not be used in a hazardous location/area where
there is danger of explosion.
Do not expose the balance to extreme moisture over long periods.
Moisture in the air can condense on the surfaces of the balance
whenever a cold balance is brought to a substantially warmer place. lf
you need to transfer the balance to a warmer area, make sure to con-
dition it for about two hours at room temperature (with the AC adapter
disconnected).
Afterwards, if you keep the balance connected to power supply, the
continuous positive difference in temperature between the inside of
the balance and the outside will practically rule out the effects of mois-
ture condensation.