Saving Automatically - wtw Oxi 330i Operating Manual

Handheld meter
Hide thumbs Also See for Oxi 330i:
Table of Contents

Advertisement

Operation
Setting the save
interval
28

4.4.2 Saving automatically

The save interval (Int 1) determines the chronological interval between auto-
matic save processes.
After the fixed interval has expired, the current data record is transmitted to
the storage and to the interface.
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <STO> key. Int 1 appears on the display.
TDS
UpH
O
S
1
n
t 1
8
O
F
Time
Baud
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
3
Set the required interval between the saving procedures with <▲>
<▼>.
4
Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.
Sal
mV/pH
% mg/l
m
S/
cm
mbar
M
cm
1/
cm
Sal
F
°
C
/
%
K
TP
°
F
Auto
Store
ARng
AR
RCL
Oxi 330i/340i
ba52306e04
03/2008