8 Creating the restoration

8.2 Design information

14
Design information
8.2
A
Screw channel
B
Titanium base
● Maintain a minimum wall thickness of 0.5 mm circularly around the
screw channel.
● The outlet point for the screw channel should not be in the area
around the contact points to the antagonist wherever possible.
● If the crown is to be reduced and then veneered immediately, make
sure that this does not narrow the screw channel. The connection
point to the adhesive base and the screw channel must not be
changed.
● Make sure that no sharp edges or corners are produced.
● The incisal/occlusal wall thickness of the ceramic above the titanium
base should be at least 1 mm.
● The circular wall thickness above the titanium base should be at
least 1 mm.
Processing instructions CEREC Zirconia meso
D3487.201.18.02.02    02.2018
Dentsply Sirona
66 21 101 D3487