The Pump Failed To Achieve Specified Performance (Failure To Reach Ultimate Vacuum); The Pump Is Noisy; The Pump Surface Temperature Is Above 100°C - Edwards E2M28 Instruction Manual

Rotary vacuum pumps
Hide thumbs Also See for E2M28:
Table of Contents

Advertisement

A373-10-880 Issue T
5.11.3
The pump failed to achieve specified performance (failure to reach ultimate
vacuum)
The measuring technique or gauge is unsuitable.
There is a leak in the external vacuum system.
The gas-ballast control is open.
The oil level is too low.
The pump is filled with the wrong type of oil.
The oil is contaminated.
The pump has not primed.
The vacuum fittings are dirty.
The inlet-filter is blocked.
The pump has not warmed up.
The motor shaft rotates in the wrong direction (three phase motors only).
5.11.4

The pump is noisy

The motor fan-cover is damaged.
The coupling and fan assembly are poorly adjusted.
The motor coupling is loose.
The motor bearings are worn.
The oil is contaminated with solid particles.
One of the pump blades is sticking.
5.11.5
The pump surface temperature is above 100°C
The ambient temperature is too high.
The cooling-air supply is insufficient or is too hot.
The cooling-air supply is blocked.
The electrical supply voltage is too high.
The exhaust filter or exhaust line is blocked.
The oil level is too low.
The pump is filled with the wrong type of oil.
The oil is contaminated.
The process gas is too hot, or the throughput is too high.
Page 30
© Edwards Limited 2014. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E2m28-ffE2m30

Table of Contents