Things needed in addition to what's in the kit:
Clean sample collection bottle (provided), clean water
(distilled, deionized, or purifi ed bottled water) for rinsing
(16oz), and soft paper towels for drying the probe.
Ensure ORP probe is securely installed (see image at
right). If not, see
Preparation before use:
Pour enough 3M KCl solution into a small cap to cover the sensors.
Soak the probe for 3-5 minutes in the solution to activate the sensor.
There is no need to activate the sensor if it is used frequently.
1
POWER ON METER
Press the
2
RINSE ORP PROBE
Rinse the ORP probe in clean water and blot dry with
soft paper towel or cloth to remove excess water.
3
SELECT MENU
Press and re-press the
test parameter (mV unit of measure next to result).
4
TEST
Fill Sample Collection Bottle with sample to be tested.
Submerge the probe in the sample and stir gently.
Then allow to sit in the bottle.
5
READ RESULTS
Record the readings after the
up. When fi nished with readings, rinse the probe with
distilled water and return it to the included storage
solution or to pH 4 buffer. NOTE: If using the eXact
iDip
the Bluetooth Test page to store the displayed result.
ORP TEST PROCEDURE
page 6
for Probe Replacement.
button to turn on the meter.
app, press the Save Result button at the top of
®
button to select the ORP
exactph.com