8
When pH
meter is turned off, press and hold
+
enter Parameter Setting. Then, press
P1-P2...P8. Press
choose parameter. Next, press
and hold
Symbol
P1
Select pH buffer standards
P2
Select automatic lock
P3
P4
P5
P6
P7
Select temperature unit
Back to factory default
P8
P1 Select standard pH buffer solution: There are two options for
standard buffer solutions, the USA series or the NIST series. Refer
to following chart:
Icons
Three-point
Calibration
©2018, Industrial Test Systems, Inc. Printed in USA
PARAMETER SETTINGS
, parameter starts to fl ash. Press
to turn off meter.
Parameter Setting
Contents
Select backlight
Temperature
compensation factor
TDS factor
Salinity unit
USA Series
1.68 and 4.00
10.01 and
to confi rm. Lastly, press
Code
USA - NIST
Off - On
Off - 1 - On
0.00 - 4.00%
0.40 - 1.00
ppt - g/L
°C - °F
No - Yes
pH Standard Buffer
Solution Series
(pH)
7.00
12.45
button to
to cycle through
Factory
Default
1 (auto)
2.00%
NIST Series
(pH)
1.68 and 4.01
6.86
9.18 and 12.45
to
USA
Off
0.71
ppt
°F
No