Remote operation
7.1 Operating via AMS Device Manager
Note
To activate simulation via AMS Device Manager or the 375 Field Communicator, simulation
must also be set to Enabled on the device. See Simulate Enable (4.12.).
● Simulation
● Simulation Value
● RAMP start
● RAMP end
● RAMP No of steps
● RAMP steplength
1. To enable simulation select Fixed value or Ramp in the Simulation field.
2. If you select Fixed value, enter a Simulation Value.
3. If you select Ramp, enter the ramp start, end, number of steps, and steplength.
4. Click Apply.
5. After simulation is complete, set Simulation to OFF and click Apply.
94
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 08/2014, A5E32221411-AC