Siemens SITRANS LR250 Operating Instructions Manual page 49

Radar transmitters
Hide thumbs Also See for SITRANS LR250:
Table of Contents

Advertisement

Entity concept:
The Entity Concept allows interconnection of intrinsically safe apparatus to associated apparatus
not specifically examined in such combination. The criteria for interconnection is that the voltage
and current which intrinsically safe apparatus can receive and remain intrinsically safe,
considering faults, must be equal to or greater than the output voltage (Uo) and output current
(Io) levels which can be delivered by the associated apparatus, considering faults and applicable
factors. In addition, the maximum unprotected capacitance (Ci) and Inductance (Li) of the
intrinsically safe apparatus, including interconnecting wiring, must be equal to or less than the
capacitance and inductance which can be safely connected to associated apparatus.
Under the entity evaluation concept, SITRANS LR250 has the following characteristics:
(input voltage) U
(input current) I
(input power) P
(internal capacitance) Ci
(internal inductance) Li
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 08/2014, A5E32221411-AC
i
i
i
5.3 Wiring setups for hazardous area installation
= 24 V
= 250 mA
= 1.2 W
= 0
= 0
Connecting
49

Advertisement

Table of Contents
loading

Table of Contents