Pressure applications
Danger of injury or poisoning.
It will be necessary to use PTFE tape or other appropriate thread sealing compound, and to
tighten the process connection beyond hand-tight. (The maximum recommended torque for
Threaded versions is 40 N-m (30 ft.lbs.) See Flanged versions (Page 37) for FEA
recommended torque values.)
Note
• On devices with a removable head, there is no limit to the number of times a device can
be rotated without damage.
• When mounting, orient the front or back of the device towards the closest vessel wall or
obstruction.
• Do not rotate the enclosure after programming and vessel calibration, otherwise an error
may occur, caused by a polarity shift of the transmit pulse.
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 08/2014, A5E32221411-AC
WARNING
Installing/mounting
4.4 Installation instructions
35