Turbidity - YSI ProDIGITAL User Manual

Professional series digital handheld meters
Table of Contents

Advertisement

3.6

Turbidity

Standards
For best results, YSI recommends the following standards for turbidity calibration:
Calibration Point
Other standards may be acceptable as long as they have been prepared according to details in Standard Methods for
the Treatment of Water and Wastewater (Section 2130 B). These standards include:
• YSI Certified AMCO-AEPA polymer-based standards (see above)
• Hach StablCal™ standards in various NTU denominations
• Dilutions of 4000 NTU formazin concentrate purchased from Hach
• Other formazin standards prepared according to the Standard Methods
The use of standards other than those mentioned above will result in calibration errors and inaccurate field readings.
It is important to use the same type of standard for all calibration points; do not mix formazin and polymer-based
standards for different points in a multi-point calibration.
When using an alternative standard (non-YSI), calibration can be completed using the following limits:
1st Calibration Point
2nd Calibration Point
3rd Calibration Point
Calibration
Standard Value
1
0 FNU [SKU: 608000]
2
12.4 FNU [SKU: 607200] or 124 FNU [SKU: 607300]
3
1010 FNU [SKU: 607400]
Min
0.0
5.0
400
Max
1.0
FNU or NTU
200
FNU or NTU
4000
FNU or NTU
Unit
37