Dissolved Oxygen - YSI ProDIGITAL User Manual

Professional series digital handheld meters
Table of Contents

Advertisement

3.5

Dissolved Oxygen

ODO calibration requires the current "true" barometric pressure. Make sure that the barometer is reading accurately
prior to ODO calibration.
Calibrating in DO% or DO% local automatically calibrates the mg/L and ppm measurement. There is no reason to cali-
brate both parameters. For both ease of use and accuracy, we recommend that you calibrate DO% or DO% Local and
not mg/L.
Figure 51 Calibrate ODO %
Calibration
ODO% and ODO% Local - Water
Saturated Air Calibration
1. Place a small amount of clean water (5 mL) in the calibration cup or
a wet sponge into the calibration sleeve (for ODO/T and ODO/CT
probes).
2. Make sure there are no water droplets on the ODO sensor cap or
temperature sensor.
3. Attach the probe guard and carefully slide into the calibration cup.
Make sure a seal is not created around the probe. Atmospheric
venting is required for accurate calibration.
4. Turn the instrument on and wait approximately 5 to 15 minutes for
the air in the storage container to be completely saturated with
water.
5. Push the
Cal
key, then select ODO. Select DO%.
6. Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
displayed in the message area.
NOTE: If you see a calibration error message, verify the barometer
reading and inspect the sensor cap. Clean and/or replace
the sensor cap as needed.
(Figure
51). "Calibration successful!" will be
35