User Manual
DOCUMENT #601119REF
®
EcoSense
pH1000A
pH, mV (ORP), and Temperature
Benchtop Meter
US ER MANUAL
E ng li sh

Advertisement

Table of Contents
loading

Summary of Contents for YSI EcoSense pH1000A

  • Page 1 User Manual DOCUMENT #601119REF ® EcoSense pH1000A pH, mV (ORP), and Temperature Benchtop Meter US ER MANUAL E ng li sh...
  • Page 2 Item #601119REF Rev C, October 2017 For the latest version of this manual, visit ysi.com ©2017 YSI Incorporated. The YSI logo and EcoSense are registered trademarks of YSI Incorporated.
  • Page 3: Table Of Contents

    C O N T E N T S Warranty ....................i Introduction ..................1 Getting Started ..................2 Initial Inspection ................2 Precautions .................2 The Electrode..............2 Instrument Power...............2 Installing the Batteries ............2 Key Pad ..................3 Main Display ................4 Instrument Operation .................5 Buffer Set Selection ..............5 pH Calibration ................5 Calibration in the pH AUTOLOCK mode ......5 Calibration in the pH NON-AUTOLOCK mode ....6...
  • Page 4 Contact Information and Service .............12 Ordering and Technical Support ...........12 Service Information ..............13 Recycling ..................13 Appendix A: Temperature coefficient of pH buffers ......14...
  • Page 5: Warranty

    Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
  • Page 6 T H I S P A G E L E F T I N T E N T I O N A L L Y B L A N K...
  • Page 7: Introduction

    I N T R O D U C T I O N Thank you for purchasing the EcoSense pH1000A, a precise laboratory benchtop meter that measures pH, mV and temperature. A built-in microprocessor stores, calculates and compensates for all parameters related to pH determinations including pH electrode temperature characteristics, electrode slope deviations and buffer solutions.
  • Page 8: Getting Started

    6 “AA” alkaline batteries. Check the label on the AC adaptor supplied with the instrument to make sure that the AC line voltage is correct. If the wrong AC adaptor is supplied, notify your YSI representative immediately.
  • Page 9: Key Pad

    KEY PAD Description Power key. Press and hold for 3 seconds to power on the meter. Pressing this key will also turn the instrument off. Up and down arrow keys are used to manually enter temperature values. They have no effect on the unit when operating in ATC mode.
  • Page 10: Main Display

    MAIN DISPLAY Number Description WAIT: This will be displayed when the unit is still waiting for a stable reading. BAT: Low battery indicator. pH: Unit and mode indicators. mV: Unit and mode indicators. ATC/MAN: ATC indicator will be displayed if a temperature probe is connected, otherwise the MAN indicator will be displayed.
  • Page 11: Instrument Operation

    I N S T R U M E N T O P E R A T I O N BUFFER SET SELECTION The pH1000A has two buffer sets: 7.00, 4.01, 10.01 pH and 6.86, 4.00, 9.18 pH. The 7.00, 4.01 and 10.01 pH buffer set is the factory default. The pH1000A will automatically recognize the buffer used as long as the correct buffer set is selected and the electrode is in good operating condition.
  • Page 12: Calibration In The Ph Non-Autolock Mode

    4. Press and hold the Stand key for 2 seconds to calibrate. The “WAIT” icon will flash until the unit detects a stable reading. The buffer used for the first point will remain lit on the display while the remaining two buffers start to flash.
  • Page 13: Viewing The Electrode Efficiency After Calibration

    A single electrode efficiency will be displayed after a single point calibration. This efficiency will always equal 100%, as the theoretical Nernst Slope will always be used for the observed slope. This is why YSI recommends at least a two point calibration.
  • Page 14: Ph Measurements

    A single electrode efficiency will also be displayed after a two point calibration. This is the efficiency between the two calibration points. Two separate efficiencies will be displayed after a three point calibration, one for the slope between 4.01/4.00 and 7.00/6.86 and another for the slope between 7.00/6.86 and 10.01/9.18.
  • Page 15: Temperature Measurements

    MEASUREMENTS ORP measurements can be made with the YSI 115-1 (item # 605376) electrode, but any ORP electrode with a BNC connection can be used. MEASUREMENTS IN THE mV AUTOLOCK MODE 1.
  • Page 16: Troubleshooting

    T R O U B L E S H O O T I N G Display Possible Cause(s) Possible Solution(s) 1. Stand was pressed before 1. Press Clear key, allow sufficient “Er1” the electrode and ATC/Temp time for the electrode and ATC/ probe settled to within +/- Temp probe to stabilize, re-press 1.5 pH of the buffer value.
  • Page 17: Specifications

    These specifications represent typical performance and are subject to change without notice. For the latest product specification information, please visit YSI’s website at ysi.com or contact YSI Tech Support. pH1000A INSTRUMENT ONLY SPECIFICATIONS These specifications reflect the range and resolution that can be displayed on the instrument, as well as the accuracy of the instrument electronics.
  • Page 18: Accessories / Part Numbers

    1725 Brannum Lane Yellow Springs, OH 45387 Web address: ysi.com When placing an order please have the following available: 1.) YSI account number (if available) 2.) Name and phone number 3.) Purchase Order or Credit Card 4.) Model Number or brief description 5.) Billing and shipping addresses...
  • Page 19: Service Information

    When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for a YSI Service Center to accept the instrument for service. The form may be downloaded from ysi.com by clicking on the ‘Support” tab, then the Product Return Form button.
  • Page 20: Appendix A: Temperature Coefficient Of Ph Buffers

    C O E F F I C I E N T O F p H B U F F E R S The temperature coefficients of pH calibration buffers 4.01, 7.00, and 10.01 (e.g. YSI buffers), as well as the 4.00, 6.86, and 9.18 buffer set, are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values.