Outputting Voltage
Example: Voltage
Output in Calibrated
Mode
Chapter 3
Will output voltage on channels 1, 2, 3, and/or 4.
–
Range is 10.922 Vdc in calibrated mode and 12 Vdc in
–
non-calibrated mode per channel.
Range can be increased to a maximum of 43.600 Vdc in calibrated
–
mode and 48 Vdc in non-calibrated mode by connecting all four
channels in series.
Remote sensing is available to ensure the selected voltage is
–
available at the load.
Figure 3-2 shows how to connect channel 1 output terminals to the load. The
D/A Converter must be physically configured to provide voltage output on
channel 1 (refer to page 17), and then instructed to output +10.00000 Vdc on
channel 1 in the calibrated mode.
Execute:
CAL1:STAT ON
VOLT1 10.00000
Figure 3-2. One Channel Voltage Output
Sets channel 1 to calibrated mode.
Configures channel 1 for voltage
and sets output at terminals to
10.00000 Vdc.
Using the HP E1328A
23