Diffusion Pump Hazards

Designers of systems utilizing diffusion pumps must design out hazards wherever possible. For hazards that cannot be
designed out, warnings, procedures, and instructions on proper use and servicing are provided. Please use guards, safety
features, and interlocks as recommended.
Refer to Table 1 for a list of general hazards and recommended actions, Table 2 on page 8 for a list of prohibited actions
that can result in explosions, and Table 3 on page 9 for a list of pressurization hazards that can result in damage to
equipment.
THE INSTALLATION, OPERATION, AND SERVICING OF DIFFUSION PUMPS INVOLVES ONE OR MORE OF THE
FOLLOWING HAZARDS, ANY ONE OF WHICH IN THE ABSENCE OF SAFE OPERATING PRACTICES AND
PRECAUTIONS, COULD POTENTIALLY RESULT IN DEATH OR SERIOUS HARM TO PERSONNEL.
Loss of utility: water and/or electricity
Overpressure in foreline
Overtemperature
Insufficient water flow through the main cooling coils
Water trapped between inlet and outlet of Quick Cool coil,
or liquid nitrogen trapped between inlet and outlet of liquid
nitrogen trap
Loss of electrical ground integrity
Positive pressure in pumping system
High voltage
Toxicity and Corrosivity
Explosion
High Throughput Diffusion Pumps
Table 1 General Hazards
Hazard
Suggested Corrective Action
Provide sufficient backup water and power supply as
necessary to effect a safe shutdown under worst case
conditions
Provide an interlock to ensure that the power supply to the
pump heater cannot be activated if the foreline pump is not
running and/or the pressure in foreline is above 0.5 Torr
(0.67 mbar)
Fit temperature sensors and pump fluid level sensors with
feedback to an interlock on the heater power supply
Use water flow sensor and feedback to interlock the heater
power supply
Provide vent or pressure relief valves for both Quick Cool
coil and liquid nitrogen trap
Incorporate ground fault interrupt circuit into heater power
supply
Integrate pressure relief valve in vacuum system to ensure
system pressure does not exceed 1 atmosphere
Prevent personnel contact with high voltages; design and
attach warnings
Vent toxic and/or corrosive gases to a safe location; ensure
adequate dilution or scrubbing to safe levels; take all action
required to meet air quality standards
Integrate pressure relief valves
Do not use hydrocarbon-based pumping fluids
7