Setting Up The Instrument; Site Requirement; Standard Delivery-Packing List - Leica HistoCore Arcadia C Instructions For Use Manual

Cold plate
Hide thumbs Also See for HistoCore Arcadia C:
Table of Contents

Advertisement

4.1

Site requirement

• Stable, vibration-free laboratory table with horizontal, flat table top, as far as possible vi-
bration-free ground.
• No direct sunlight or strong temperature fluctuations. Room temperature consistently be-
tween +20 °C and +30 °C.
• Relative air humidity maximum 80 %, non-condensing.
• The instrument should be set up in such a way that the air circulation is not impaired.
• The instrument must be installed in a place that ensures an easy disconnection from the
power supply. The power cable must be in a place that can be easily reached.
At a room temperature of > +30 °C, the working temperature of the cold plate of -6°C
may not be reached at all points.
To ensure proper function and an easy disconnection of the power cable from the in-
strument, there must be gap of at least 15 cm behind the instrument. Failure to observe
this distance may result in serious damage to the refrigeration unit of the device. The
instrument should not be operated in hazardous locations.
4.2
Standard delivery—packing list
Qty
Designation
1
Basic unit HistoCore Arcadia C,
220-240 VAC
220-240 VAC, China
110-120 VAC
100 VAC
220-240 VAC
220-240 VAC, China
110-120 VAC
100 VAC
4
Sets of spare fuses: 220-240 VAC, 5A 250 V
1
Instructions for Use (printed German/English with language CD 14 0393 80200)
HistoCore Arcadia C – Cold Plate
100-120 VAC, 10A 250V
4.

Setting up the instrument

Order No.
14 0393 53599
14 0393 54102
14 0393 54092
14 0393 54091
14 0393 57262
14 0393 57263
14 0393 57261
14 0393 57260
14 6000 05015
14 6000 05078
14 0393 80001
13