Setting up the instrument

4.1

Site requirement

• Stable, vibration-free laboratory table with horizontal, flat table top, as far as possible vi-
bration-free ground.
• The instrument must not be placed in the vicinity of the air outlet of an air conditioner and
must be shielded from strong sunlight (window).
• To ensure a fully functional heat sink, there must be gap of at least 15 cm behind the
instrument.
• The instrument must be installed in a place that ensures an easy disconnection from the
power supply. The power cable must be in a place that can be easily reached.
• The vicinity of the work area must be free of oil and chemical vapors.
The installation location must be well-ventilated and free of ignition sources of any
kind. The instrument should not be operated in hazardous locations.
16
Instructions for Use V 1.7 Rev. J – 12/2017