Warnings (Continued)
Safety instructions - Working with the instrument
Paraffin is flammable and should therefore be handled with due care.
Do not use sharp tools to remove solidified paraffin from the work areas, as this may
destroy the coating on the surface. Use the plastic spatula supplied with the instrument.
During operation, the paraffin tank, embedding mold tray, cassette tray, work area as
well as the forceps holder are hot.
Risk of burning!
Do not move the instrument during operation.
Do not store any combustible and flammable substances near the instrument. There is
a fire hazard if work with an exposed flame (e.g. Bunsen burner) is carried out in the
direct vicinity of the instrument (solvent vapors). Therefore, keep all ignition sources at
least 2 meters away from the instrument!
Please allow 30 minutes before touching the instrument after shutting down.
Failure to observe the instructions specified by the manufacturer may result in damage
to the protection provided by the instrument.
Hazards - Servicing and cleaning
Switch off the instrument each time before servicing and pull out the power plug.
When using cleaners, please comply with the safety instructions of the manufacturer
and the laboratory safety regulations.
Before changing defective fuses, the instrument has to be disconnected from the power
supply. Fuses in the fuse holder on the back panel can be replaced by the user.
Liquid must not get into the instrument while operating or cleaning it.
2.3

Integrated safety devices

The instrument is equipped with the following safety features and devices:
Fuses in the heating elements
All of the instrument's AC heating elements are equipped with overheating fuses, which trip if
an AC heating element overheats and switches the element off.
Fuses will only automatically reset when the instrument is disconnected from AC power
and the heating element temperature drops below 50 °C.
Note the only way the user can disconnect the instrument from a power source is to
remove the power plug.