PATIENT MESSAGES

Various messages may be displayed during your therapy; some include an audible alert
that is controlled by Master Volume Setting.

1. Notifications

These are friendly reminders to clean or replace components and, if enabled, are
generated as follows. Dismiss from Display by Button Press or Knob Movement.
Displayed Notification
Air Intake Filter Due for Cleaning
Mask, Tubing, and Chamber Due for Cleaning

2. Messages

These identify conditions that may require action by you/your provider or are
informational only. Dismiss from Display by Button Press or Knob Movement.
Displayed Message
Mask Off Please Check Mask Fit
Auto-STOP Detected
Auto-START Detected
Wake-Up Alert <current time>
Error with Code, Prescription Not
Updated
Card Detected Please Wait ...
Card Removed Send to Provider
Card Error Contact Provider
Settings Not Updated Contact
Provider
Card Transfer Successful
Card Detected Ready for use
Firmware Update Please Wait ...
EN - 48
Recurrence interval
50 hours of use (while breathing)
35 hours of use (while breathing)
Condition for Occurrence
10 seconds after mask off condition detected.
When Auto-STOP occurs.
When Auto-START occurs.
When Wake-Up Alert activates (alarm clock).
When SmartCode Rx is entered but incorrect.
When SD card is detected. Do not remove SD
card while this message is displayed.
When SD card is removed after data transfer.
If SD card error condition detected.
When prescription change via SD Card fails.
When valid SD Card inserted and data transfer
complete.
When valid SD Card is inserted.
When SD Card containing firmware update is
inserted.

3. Add-On Device Messages

These messages may be displayed when Add-On items are in use, Dismiss from Display
by Button Press or Knob Movement.
Displayed Message
Oximeter Logging Started
SpO2 _____
Pulse Rate _____
Oximeter Logging Stopped
Oximeter Finger Probe Off
Oximeter Finger Probe On
SpO2 _____
Pulse Rate _____
Wireless Modem Low Battery, Please
Recharge

4. Service Codes

These are divided into Critical or Non-Critical. Critical service codes put the unit in a fail-
safe (blower off) state.
Displayed Message
Critical Service Code E0X - Where (X) is a
number or letter
NOTE – Code E01 is always shown in
English since language setting is unknown
Service Code E8X (non-critical - Where (X)
is a number or letter
NOTE- Service Code alerts are unaffected by notification enable/disable setting.
Condition for Occurrence
When Oximeter first detected and logging
starts
When Oximeter signal is lost and logging
stops
When Oximeter finger probe off condition
detected
When Oximeter finger probe off condition
corrected
When DV6WM is paired and battery level is
low
Action
Contact your equipment provider to
return unit for service. Message remains
on Display and cannot be dismissed.
Contact your equipment provider.
Dismiss from Display by Button Press or
Knob Movement.