Patient Messages; Reminders; Notifications - DeVilbiss DV54 AutoAdjust CPAP Series Owner's Manual

Positive airway pressure device
Table of Contents

Advertisement

Available languages

Available languages

pAtient messAges

PaTienT messaGes
Your IntelliPAP will alert you of issues that may require some action on your part in order to
make your therapy more effective. These messages are displayed when the device is turned
on, either by pressing the ON/OFF button or by breathing into the mask. You can press any
key or button to clear the message from the display, but the IntelliPAP will operate normally
even if the message is not cleared.
Two kinds of messages are displayed: Reminders and Notifications.

Reminders

Reminders alert you when parts of your system need to be replaced. Some components of
your system wear over time and, if not replaced, may compromise your therapy. If you see
a message on the display that is not listed below, you may have the optional DeVilbiss
SmartLink™ Module attached to your device, which provides additional messages. Refer to
the documentation that was provided with the SmartLink Module.
Clean Filter – The filter should be checked every 10 days for signs
of dirt or wear so that it can be cleaned as needed. This message
helps to remind you to check it regularly.

Notifications

Notifications identify conditions in your IntelliPAP that may require action by you or your
home care provider in order to maintain a high level of therapy. If you see a message on the
display that is not listed below, you may have the optional DeVilbiss SmartLink Module
attached to your device, which provides additional messages. Refer to the documentation that
was provided with the SmartLink Module.
NOTE
-The values displayed are examples only.
Delay Running xx Minutes Left – While a comfort delay is active,
this message is flashed on the display every 5 seconds to let you
know how much time is left in the delay.
Mask Leak – This message means the IntelliPAP has detected a
large amount of airflow during therapy for at least 10% of the time
during your previous use session. This notification is displayed when
the device is turned on. If this message is displayed, put on the mask and adjust the headgear
to ensure the mask is properly fitted to your face. Follow the mask manufacturer's directions
for adjusting the mask and headgear for proper fitting. Available on AutoAdjust models.
Mask Off – This message appears whenever a high airflow
condition occurs. High airflow is normally due to a poor mask fit or
a removed mask. Check for leak around the mask seal and make
adjustments as necessary according to the mask manufacturer's instructions. This message
will be displayed until the high airflow problem is corrected. If the high airflow condition
persists for about 20 seconds, the device will automatically turn off (if AutoOFF is enabled).
Device Fault – If a device fault message is displayed, refer to
Troubleshooting for instructions.
EN
Clean Filter
Delay Running
10 Minutes Left
Mask leak =
25 % Check Fit.
Mask Off. Please
check mask fit.
Device Fault E07
Call Provider
A-DV54
15