Installation
5
Model 2-Wire
5.4.4 mA Multiple Loops
The following figure is an example of a multiple loop system. Two loops are operated from a single power supply.
An arrangement of this type is suitable for measuring temperatures at two or more stations with an independent
readout for each station. The advantage is the economy of a single power supply for all loops.
An important consideration in this system is the current capacity of the power supply. For example, if both loops
are measuring full-scale temperature, the total supply current is calculated as follows:
=
+
= 20 + 20 = 40
1
2
Figure 5-6: Principle Circuit Diagram: Infrared Sensor with Multiple Loads
Controllers
Infrared Sensors
Indicators
I
Loop1
I
Loop2
Power Supply
I
Loop
5.4.5 Alarm Output AL
The maximum current carrying capacity for the alarm output is 150 mA. Use the circuit diagram below. The alarm
output AL of the instrument is not electrically isolated from the power supply.
Figure 5-7: Exemplary Wiring the Alarm Output AL for the 2-Wire Sensor
35
1.800.561.8187
information@itm.com
www.
.com