The table below lists some of the problems you may experience with your device or mask and pos-
sible solutions to those problems.
P
ROBLEM
Nothing happens when
you apply power to the
device. The audible in-
dicator does not sound
and the backlights on
the buttons do not light.
The device does not
operate when you press
the
button. The air-
flow does not turn on.
5-4
W
I
H
HY
T
APPENED
There's no power
If you are using AC power, check the outlet
at the outlet or
power and verify that the device is properly
the device is un-
plugged in. Make sure the AC power cord is
plugged.
connected correctly to the power supply and
the power supply cord is securely connected to
the device's power inlet. If the problem contin-
ues to occur, contact your home care provider.
Return both the device and power supply to
your provider, so they can determine if the
problem is with the device or power supply.
If you are using DC power, make sure your DC
power cord and battery adaptor cable con-
nections are secure. Check your battery. It may
need recharged or replaced. If the problem
persists, check the DC cord's fuse following the
instructions supplied with your DC cord. The
fuse may need to be replaced. If the problem
still occurs, contact your home care provider.
If you are using a humidifier, make sure you fol-
low the instructions for applying power that are
provided with your humidifier.
There may be a
Make sure the device is powered correctly, fol-
problem with the
lowing the instructions on the previous page.
blower.
If the audible indicator sounds and the button
backlights turn on when you apply power, but
the airflow does not turn on, there may be a
problem with your device. Contact your home
care provider for assistance.
Note: When the device is functioning cor-
W
D
HAT TO
O
rectly, after you press the
the device beeps and the airflow turns
on after a slight delay. This brief delay
is normal.
REM
P
M S
STAR
RO
ERIES
button,
U
M
SER
ANUAL