P
RESSURE
Pressure Increments:
4.0 to 20.0 cm H
O (in 0.5 cm H
O increments)
2
2
Pressure Stability:
<10.0 cm H
O (±0.5 cm H
O)
2
2
≥10.0 to 20.0 cm H
O (±1.0 cm H
O)
2
2
Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15,
and 20 BPM @ 20° C (±5° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
Maximum Flow: 35 LPM
Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15,
and 20 BPM @ 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
D
ISPOSAL
Dispose of the device in accordance with local regulations.