Connection To Various Breathing Systems - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Proper use of these products depends on careful reading and
understanding of labeling and instructions. Please refer to the
INOmax DS
and INOblender operation manuals for guidance. Also
IR
for use.
INOmax DS
Warnings:
IR
• INOmax DS
subtracts gas from the breathing circuit via the
IR
some ventilators. The trigger sensitivity of the ventilator should
be checked after connecting the INOmax DS
circuit.
• Patient disconnect and high-pressure alarms are required for
the ventilator.
INOmax DS
Cautions:
IR
• Insert the Injector Module on the dry side of the breathing circuit
• Avoid medications interfering with the gas monitoring system;
administer any aerosolized medications distal to the sampling tee.
Circuit Connection
Diagrams
to the breathing
IR
17
Part No. 20751 Rev-01
2014-08