2. Working Conditions
10°C to 30°C / 15% up to 80% humidity
3.
Electrical Power Supply
Either by batteries or current power supply
Power supply:
-
Connection by power adapter
-
Batteries (6 x AA batteries
4.
Technical Data
-
Accuracy: ± 0,5 % of Capacity
-
Data Sampling Rate: 1.000 Hz
Model
Capacity
FK 10T
FK 25T
FK 50T
FK 100T
FK 250T
5.
Operation
a.
Display
Measuring Result
Measuring Units
Measuring Direction
Indication of PEAK Mode
b.
Operating keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
MAX:
Changes between Track and Peak mode (capture of
maximum values)
UNIT:
Press shortly:
N, kg, lb, ou
FK_T-BA-e-1110
Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: info@sauter.eu
Instruction Manual
FK Tensio
)
Resolution
10 N
0,005 N
20 N
0,010 N
50 N
0,020 N
100 N
0,050 N
200 N
0,100 N
(4)
(3)
(1)
(2)
Select unit:
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
ZERO:
Zeros the measuring result (Tara function)
Cleans the peak value (in Peak mode)
c.
Display Return
When the instrument itself is turned that the head faces
down, the display returns automatically.
d.
Measurement (Track Modus)
Display (1) shows the continuous force in a defined
direction (3)
To zero the display, press ZERO:
e.
Peak-Hold Function (Peak Mode)
Please press: MAX
6.
Backside Fixture to a test stand
Fixture by 4 x M3 screws.
7.
Adjusting procedure
Observe stable environmental conditions. A short warming
up time of approx. 1 minute is recommended for
stabilization.
Switch on the unit by pressing the Unit-key and keep it
pressed until in the display first appears shortly "CAL".
After that there appears the fix size of the adjustment
weight. Attach the adjustment weight, a short time later
appears "F". Then the instrument returns automatically to
the measuring mode.
In case of an adjustment error or incorrect adjusting weight
the display will show "E"; repeat adjustment process.
8. CE Declaration of Conformity