5.7.3 Non-verifiable Models
FOB 1.5K0.5, FOB 3K1, FOB 6K2, FOB 7.5K0.5L, FOB 15K1L, FOB 30K2L
Turn balance off and hold
erate the
-key 3x times during this test procedure.
Now let go of both keys. In the display appears "CAL". Press key
play appears "Load".
Place the tare weight (nominal value, see "Techn. Specifications" chapter 1) in
the centre of the weighing pan, then press the key
the display for approx. 2 seconds, followed by the nominal value of the adjust-
ing weight.
Remove the adjusting weight from the weighing plate to complete adjusting.
5.7.4 Check software settings for verification of a balance
(only models FOB 1K1M, FOB 2K2M, FOB 5K5M):
For the adjustment function, the balance must be switched over to a service
mode. This service mode is activated by actuating the adjustment switch
twice.
In the service mode all parameters of the balance can be modified. There is a
special setting named „TRAD".
The service parameters may not be modified, as this could damage the bal-
ance settings.
Especially the setting „TRAD" is very important for a verification. This setting
activates and deactivates the verifiable settings of the balance.
Prior to a verification the setting „TRAD" must be revised.
Press the
Press
button repeatedly. The setting „Trad" must not appear (see graph-
ic).
FOB-BA-e-1619
key, test procedure appears on the display, op-
key and keep it pressed until „A_OFF" appears on the display.
in dis-
. "CAL" will appear on