MODEL CLARITY II
5.5.3 Procedure: Selecting Turbidity or TSS
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
Outputs
M M e e a a s s u u r r e e m m e e n n t t
Configure?
S S e e n n s s o o r r 1 1
Sensor2
SensorN is for:?
T T u u r r b b i i d d i i t t y y
Units?
N N T T U U
FTU
Units?
p p p p m m
mg/L
S S i i g g n n a a l l A A v v e e r r a a g g i i n n g g
Bubble Rejection
Signal Averaging?
Signal Averaging
B B u u b b b b l l e e r r e e j j e e c c t t i i o o n n
Bubble Rejection?
O O n n
28
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
Hold
2. Choose Measurement.
Alarms
>>
3. Choose Sensor 1 or Sensor 2. For a single input configuration, the
Sensor 1 Sensor 2 screen does not appear.
4. Choose Turbidity or TSS.
TSS
5. Choose the desired units:
FNU
a. For turbidity choose NTU, FTU, or FNU.
b. For TSS choose ppm, mg/L, or none.
none
6. Choose Signal Averaging.
7. Choose the desired signal averaging. For most applications, 20 sec is
suitable.
2 2 0 0
sec
8. The display returns to the screen in step 6. Choose Bubble Rejection.
9. Choose On to enable bubble rejection software. Choose Off to disable.
Off
10. Press EXIT to return to the previous screen. To return to the main display,
press MENU followed by EXIT.
SECTION 5.0
PROGRAMMING THE ANALYZER