QUICK START GUIDE
1. Refer to Section 2.0 for installation instructions.
2. The sensor cable is pre-wired to a plug that inserts into a receiving socket in the analyzer. The cable also passes
through a strain relief fitting. To install the cable...
a. Remove the wrenching nut from the strain relief fitting.
b. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
c. Slide the wrenching nut over the plug and screw it onto the fitting.
d. Loosen the cable nut so the cable slides easily.
e. Insert the plug into the appropriate receptacle on the circuit board.
f.
Adjust the cable slack in the enclosure and tighten the cable nut. For the wall/pipe mount version, be sure to
leave sufficient cable in the enclosure to avoid stress on the cable and connections.
g. Plug the cable into the back of the sensor.
h. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first applied to the analyzer.
3. Make power, alarm, and output connections as shown in the drawing below.
Panel Mount version
The release clip on the sensor plug
faces the top of the analyzer enclosure.
FOR CLARITY II TURBIDIMETER
Power, Alarm, and Output Connections
Wall/Pipe Mount version
The release clip on the sensor
plug faces the user.
CONTINUED ON THE FOLLOWING PAGE