MODEL CLARITY II
5.4.3 Procedure: Configuring Alarms
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
Outputs
Measurement
Alarm Setpoints
A A l l a a r r m m C C o o n n f f i i g g u u r r e e
Disable all
Alarms?
Y
Alarm Config?
A A L L 1 1
AL2
AL1 is for?
S S e e n n s s o o r r 1 1
Sensor2
AL3 is for?
Sensor1
Sensor2
24
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
Hold
2. Choose Alarms.
A A l l a a r r m m s s
>>
3. Choose Alarm Configure.
4. To disable all alarms, choose Y (yes). The display returns to the screen
shown in step 2. If you want to use the alarm relays, choose N. Go to step 5.
N N
5. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
AL3
6. For AL1 or AL2
a. Choose Sensor 1 or Sensor 2. For a single input configuration, the
Sensor 1 Sensor 2 screen does not appear.
b. Choose High or Low.
c. Set the alarm Deadband.
7. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
F F a a u u l l t t
8. For AL3
a. Choose Sensor1, Sensor2, or Fault.
b. For Sensor1, select High or Low and set the deadband.
c. Choosing Fault means AL3 will activate when a sensor or analyzer
fault exists. There is no user setting to make.
9. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
SECTION 5.0
PROGRAMMING THE ANALYZER