Automatic Temperature Data Output - Siemens Ardoptix PT 140 AF4 Operating Manual

Pt 14 series two-colour pyrometer
Table of Contents

Advertisement

Operating manual
s
13: Quotient ATD Trigger
14: Lambda 1 ATD tAct
15: Lambda 2 ATD tAct
16: Quotient ATD tAct
------------------------------------------------------------------
Your choice>
In the submenu "Options" you can program a keylock. If keylock is acti-
P100
vated and you wish to change a parameter setting, enter Code
for full access. If you enter an incorrect code, you can view the configura-
tion but you will not be able to change the setting.
------------------------------------------------------------------
Submenu OPTIONS
------------------------------------------------------------------
Autoprint ......... on (cyclic)
Print cycle time .....
0.1 s
Protocol address .......... 001
Display ........... temperature
Key lock .................. off
Unit .................. Celsius
A: Set autoprint function
T: Set output cycle time
P: Set protocol-address
D: Set display function
E: Set key lock
F: Set unit Celsius/Fahrenheit
R: Switch off pyrometer
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>I/O >OPTIONS >

19.3.6 Automatic temperature data output

For continuous serial transmission of temperature data, you must acti-
vate automatic temperature data output.
In the submenu "Options", command "A" activates (on) / deactivates
(off) this function.
If ATD is activated, the temperature data will only be transmitted after
successful completion of measurement.
If ATD is deactivated, press command "T" to select the cycle time in
which the current temperature data are transmitted through the serial in-
terface.
When automatic temperature data output is activated, the pyrometer will
not display the parameter settings when it is switched on. Instead, the
pyrometer will instantly begin transmitting data.
35