Treatment Screen Parameters - ResMed H5i Clinical Manual

S9 series vpap adapt positive airway pressure device heated humidifier
Hide thumbs Also See for H5i:
Table of Contents

Advertisement

Treatment screen parameters

Parameter
Treatment screen
CPAP
ASV
Ramp
Oxygen saturation
(SpO
)*
2
Leak
Minute Ventilation (MV)
Target minute
ventilation (TgMV)
Pulse*
Respiratory rate (RR)
Tidal volume (Vt)
Pressure support (PS)
* Only available via the oximeter adapter. .
14
Modes
Description
CPAP
ASV
Shows the fixed treatment pressure.
ü
Shows the treatment pressures.
ü
Orange icon shows that device is ramping up.
ü
Measure of the saturation of blood hemoglobin with oxygen,
ü
ü
expressed as a percentage (sampled every second).
Estimate of the total rate of air escaping due to mouth and
ü
ü
unintentional mask leaks, expressed in L/min (5-breath moving
average).
Volume of air breathed in, or out within any 60-second period,
ü
ü
expressed in L/min (5-breath moving average).
Minute ventilation the device is attempting to achieve. Pressure
ü
support increases if the minute ventilation falls below the target,
and decreases if it goes above the target.
Number of heart beats in a 60-second time frame (sampled
ü
ü
every second).
Frequency of breathing, expressed as the number of breaths
ü
ü
per minute (5-breath moving average).
Volume of air inspired or expired in one respiratory cycle
ü
ü
(breath), expressed in mL (5-breath moving average).
Difference between the peak pressure at the end of inspiration
ü
and the minimum pressure at the end of expiration.

Hide quick links: